The Nuclear Power Industry trusts Tavis for some of their critical air flow controls, with very low differential pressure transducers that have been seismically qualified. So do oil companies measuring high, down-hole pressures at hot temperatures. Conventional power companies also call on Tavis for their special sensing needs. When a process requires a rugged, bulletproof, install-it-and-forget-it measurement that the cheap sensors can’t handle, a Tavis transducer or transmitter is the right sensor investment. This is especially true for applications that must monitor very low delta pressures.


Power and Industry customers have looked to Tavis for custom sensors that offer:
  • Integral radiation shielding
  • Internal heater-controlled for added accuracy
  • Either 2- and 3-wire current loop (4-20 mA) outputs or 4-wire voltage outputs
  • 110 VAC operation
  • Very Low Delta Pressure, some with high line and overpressure capability
  • Fully sealed environmental enclosures or all-welded designs
  • Seismic Qualification
  • Acoustic tolerance
Unit of Measure

General

Pressure Range1 N/A 0 - 0.1 inch H20D to 0 - 15 inch H20D

Line Pressure

N/A 20 psia MAX

Proof Pressure

N/A 1 psid or 200% FS, whichever is greater

Burst Pressure

N/A >>200% FS

Sensor Type

N/A Variable Reluctance

Weight

N/A 12 lbs Max

Wetted Materials

N/A 410 SST
316 CRES
Inconel 600/625

Construction

N/A 2024 Aluminum Case with Polyester Epoxy Paint Internal Heater 110 VAC Input to Internal Terminal Block (Pigtail installed by User) Zero and Span Adjust Pots

Fittings / Ports

N/A Female 3/8 - 18 NPT Fittings

Electrical

Input Voltage

N/A 24 - 32 VDC (Transducer) 110 VAC (Heater)

Output Signal

N/A 4 - 20 mA

Power/Current

N/A 17 - 20 mA (Transducer) 50 W (Heater)

Electrical Interface

N/A 110 VAC and 28 VDC Pigtail to Internal Terminal Strip (Pigtails installed by User)

Reverse Polarity Protection

N/A No

Insulation

N/A >100 MΩ @ 100 VDC

Output Impedance

N/A 0 - 500 Ω Load Impedance

Environmental

Compensated Temps

N/A 0 °F to +130 °F (Internal Heater Controlled)

Shock

N/A 100 grms 11 ms

App Media

N/A Air Gases

Performance

Thermal Error

N/A Heater Controlled

Total Error Band (TEB)

N/A ±0.5% FSO Max
  • 1 The designs above are intended to convey just a few capabilities previously used in industrial settings for discussion purposes. Tavis specializes in custom configurations and capabilities for your specific application. Our engineers will be happy to assist you with your unique measurement requirements.